Table Top Height Adjustment

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Non-ferrous standard head screwdriver1--

About this task

Overview

This procedure provides instructions on how to adjust the height of the table top.

Procedure

  1. Dock the patient table.
  2. Use the UP foot pedal to raise the patient table to full up position, and move the cradle completely into the magnet bore to expose the table top height adjustment screw.
    Figure 1. Table top height adjustment screw location
    1Table top height adjustment screw
    2Front of table
  3. Use a standard head screw driver to remove the cover plug.
  4. Adjust the screw to set the table top height equal to the bridge height.
    Figure 2. Rolling surface on bridge and table
    Item Description
    1Cradle
    2Rolling surface (path) on left and right sides of bridge and table top. Straight edges are used to show the paths.
    3Bridge
    4Table
  5. Replace the cover plug.

Finalization

  1. Dock the table.
  2. Confirm that the cradle release functions properly.
  3. Confirm that the LPCA connects to the cradle and that the cradle can move in and out of the bore.