• Object ID: 00000018WIA30EADD20GYZ
  • Topic ID: id_13107501 Version: 4.0
  • Date: Mar 23, 2020 2:45:30 PM

1.5T Breast Array Coil Setup for MCQA or SNR Test

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable30 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
10 cm Spherical Unified Phantom2Unified phantom set

46-317586G1

-
Unified Phantom Positioner1Unified phantom set

5343432

-
Table 3. Required conditions
ConditionReferenceEffectivity

For HDx, HDxt, and HDe, install the following coil configuration names to run this tool: HDBreastServ, HDBreastRtServ, and HDBreastLtServe.

--

For Discovery MR450 and Optima MR450w, use the following pre-installed coil configuration names to run this tool: HDBreastRight and HDBreastLeft.

--

About this task

Follow this procedure to prepare for the automated SNR test using the 1.5T HD Vibrant Breast Array Coil, GE/USAI catalog number M3087JG or M7000FZ.

Note:

Coils do not ship with phantoms. Phantoms come in a unified phantom set with the MR system.

Unified Phantom Procedure

Procedure

  1. Install the flat filler panel on the GEM table.
    Figure 1. Flat Filler Panel on GEM Table
  2. Place the coil on the patient table, and insert the phantom positioner into the coil as shown below.
    Figure 2. Phantom Positioner Inserted in Coil

    The position of the spherical phantoms from the side of the coil is shown below.

    Figure 3. Position of Spherical Phantoms
  3. Place the phantoms on the phantom positioner as shown below, connect the coil to Port A of the LPCA, landmark on the center of the bridge between both chambers, and Advance to Scan.
    Figure 4. Landmark on Center of Bridge
  4. Perform MCQA Test, if available. Otherwise, perform Surface Coils SNR Test.

What to do next

Finalization

No finalization steps.