• Optima MR450w BASE 1.5T System Service Methods
  • 5690012-2EN Revision 3
  • Object ID: 00000018WIA30C6AE20GYZ
  • Topic ID: id_13107546 Version: 3.0
  • Date: Aug 29, 2019 1:34:43 AM

Front Cover Removal and Installation

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes for front cover; 30 minutes for GEM light strips15 minutes
Tools and test equipment
ItemQuantityPart numberManufacturer
Pair: Cut-Resistant Gloves1--
Nonmagnetic Titanium Service Tool Kit, Large Set15112581-
EPX Plus II Applicator15265376-
Mixing Nozzle, 3M Scotch-Weld EPX Applicator, 50 ml, PN 62-9154-9148-4 or Equivalent146-252065P154-
Consumables
ItemQuantityPart numberManufacturer
Plexus MA300 Adhesive, 50 ml15774729-
Adhesive Cable Tie Mount146-208747P1-
DV Light Assembly15766238-
5.6 inch Cable Tie146-208758P2-
Replacement parts
ItemQuantityPart numberManufacturer
Front Cover2See FRU manual-
Table 1. Safety

About this task

Overview

This document contains the procedures for front cover removal and installation. The front covers consist of two halves and contain the emergency stop and control panels for the system.

Preliminary

Procedure

  1. Turn off and perform LOTO on the RF and PEN cabinets. See the MR Service Safety Manual, PN 5452735.
  2. Undock the patient table and position the table away from the magnet.

Front Cover Removal

About this task

Figure 1. In-Room Display (IRD) Front Covers

Procedure

  1. Remove the side cover.
  2. Remove the skirt cover.
  3. For an in-room display (IRD) system:
    1. Remove the alignment laser light cover by unscrewing the two screws. The cover is held in place by a popper on the lower left corner.
      Figure 2. Removing the Laser Light Cover
    2. Remove the two trim rings by pulling away from the enclosure. They are held in place by four poppers on each side.
      Figure 3. Removing the Trim Ring
  4. For a 7-segment display (7SD) system:
    1. Remove the left trim ring.
    2. Remove the one screw holding the 7-segment display bezel.
    3. Remove the right trim ring.
      Figure 4. 7-Segment Display Bezel
  5. For both systems, remove the two M6 screws that secure the front bridge cover.
    Figure 5. Front Bridge Cover Screws
    Figure 6. Front Bridge Cover Removed
  6. Remove the front cover(s) by unscrewing the seven screws on the left-hand front cover and the six screws on the right-hand front cover. Unscrew the two additional screws shared by both sides. There are fifteen screws total.
    Note:

    Be careful of the operator control panel connection cables.

    Figure 7. Unbolting Front Cover
  7. To free each front cover, disconnect the cables that connect the control panel to the magnet Replacing Operator Display and Control Panels and disconnect the room ambient temperature sensor cable from the left front cover Room Ambient and Bore Temperature Sensor Replacement.
    Figure 8. Left Front Cover
  8. Reverse the steps to install the replacement front cover.

Vertical Light Strip Replacement

About this task

For left or right light strip removal, remove the corresponding side middle and side front covers.

Procedure

  1. Verify that LOTO is still in place and the patient table is undocked (see Preliminary).
  2. Remove the corresponding side covers. See Side and Top Cover Removal and Installation.
  3. Disconnect the power cord from the corresponding light strip.
    Figure 9. Disconnect the light strip power cord
  4. Remove the translucent light cover.
    Figure 10. Remove the translucent light cover
  5. Remove the clips holding the lights to the cover. Cut the wire ties on the back side holding the light strip wire and remove the light strip.
    Figure 11. Remove the clips
    Figure 12. Cut the wire tie holding the strip wire
    Note: If the current LEDs in the cover are already mounted to adhesive mounts, proceed to Step 10.
  6. CAUTION: Nitrile gloves must be worn when using alcohol wipes and glue.
    Assemble the glue gun applicator using the 1:1 plunger, following the manufacturer's instructions included in the box. Make sure to connect the glue and mixing nozzle to the applicator assembly.
    Figure 13. Glue gun with 1:1 plunger
    Figure 14. Assembled glue gun
  7. Clean the location where the adhesive mounts will be installed using the alcohol wipes provided in the FRU kit.
    Figure 15. Clean the LED groove
  8. Install the adhesive mounts into the LED groove. Make sure to install one at the top and one at the bottom of the cover, with the remaining 9 spaced approximately 200 mm apart in the LED groove. Placement is not critical, but should span the whole groove.
    Figure 16. Adhesive mount installation and placement
  9. Using the gun applicator, liberally apply glue to the vertical sides of the adhesive mounts. Wait until the glue is no longer tacky (approximately 10 minutes) before continuing to the next step.
    Note: It is critical that glue coverage does not extend outside of the LED groove, and that glue does not interfere with the slots for cable tie installation.
    Figure 17. Glue application
  10. Install the LED strip starting from the bottom to the top of the panel, pealing the backing off the LED strip (approximately 12 inches) to expose the adhesive tape, and continuing up the side as the LED strip is placed. Between adhesive mounts, make sure to press the LED onto the cover to eliminate any gaps.
    Figure 18. LED installation
  11. Loop the cable ties through the adhesive mounts, locking the LEDs in place.
    Figure 19. Cable tie installation
  12. Clip the excess from the cable ties.
    Figure 20. Clip excess cable tie
    Figure 21. Proper LED placement
  13. Replace and reconnect the light strip by reversing the above instructions. Ensure proper fit.

Finalization

Procedure

  1. For in room display (IRD) systems:
    1. Whenever the IRD or trackball control panels are disconnected and then reconnected, the host PC needs to be rebooted.
    2. After reboot, from the Common Service Desktop, select Diagnostics > Hardware Location > Magnet Room > In Room Display.
    3. The diagnostic displays the current status of the IRD and trackball connectivity. If the status is not installed, recheck connections, reboot, and check the In Room Display diagnostic again.
  2. For 7-segment display (7SD) systems, check the functionality of the item replaced. Run the associated diagnostic, if necessary. From the Common Service Desktop, select Diagnostics > Hardware Location > Magnet Room > SRI Functional Tests.