Power Harness Replacement
Prerequisites
Overview
Procedure
- Remove the following gantry covers: side, top and front cover.
-
Side covers.
-
Top covers.
-
Front cover.
-
danger- Turn OFF all 3 switches (Axial Drive, HVDC, 120 VAC) on the STC backplane. Perform Lockout Tagout prior to continuing.
danger- Position the gantry with the XRT at 6 o'clock.
- Remove the top cover from the Collimator by removing the five pan head 4x8mm mounting screws with spring washers.
- Cut the ty-raps holding down the Power Harness.
- Disconnect the Cam A motor drive power connector.
- Disconnect the Cam B motor drive power connector.
- Disconnect J12 from CCB.
- Remove Power Harness and place in packaging to return.
- Place the new Power Harness on the frame.
- Connect J12 to the CCB.
- Connect the Cam A motor drive power connector.
- Connect the Cam B motor drive power connector.
- Ty-Rap the Power Harness.
- Replace the top cover of the Collimator using the five pan head mounting screws with spring washers and tighten.
Finalization
- Perform CAM A,B Amplifier Check.
- Perform CAM A,B Encoder Check.