- Topic ID: id_11038790
- Version: 5.0
- Date: Jan 20, 2020 8:35:34 PM
Reset TnT
Prerequisites
Overview
This document defines the procedures to reset the Generator Tracking and Trending data and filament aging calibration to factory values.
Procedure
- Select Service Desktop.
- Select Hardware ResetScan.note:
This step guarantees Jedi is in Applications mode and Reset TnT will work properly.
- Select DIAGNOSTICS from Service Desktop.
- Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
- Click RESET TNT DATA
- Select (highlight) TUBE TNT DATA and GENERATOR TNT DATA .
- Click RESET.
- Click OK in Attention window.
- Click CANCEL in Reset TnT Data window.note:
If process fails, take appropriate corrective action and repeat procedure. If the process fails persistently, perform Initialize Generator Database.
- Click DISMISS in Generator Tool - JEDI window.
Finalization
- Filament Calibration
- Save System State.