- Topic ID: id_11038924
- Version: 5.0
- Date: Jan 20, 2020 8:32:31 PM
Initialize Generator Database
Prerequisites
This document defines the procedures to reset the generator database to factory values.
Procedure
- notice
- Select DIAGNOSTICS from Service Desktop.
- Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
- Click INITIALIZE DATABASE.
- Click OK in Attention window.
- Click OK in Initialize Database window.
![]() |
Finalization
- Restore Generator Runtime Parameters (See Save Restore
Generator Runtime Parameters)
note:
If Save Generator Runtime Parameters failed, perform Reset TnT, Filament Calibration.
- Click DISMISS in Generator Tool - JEDI window.
- Save System State.