Reset TnT

Prerequisites

Overview

This document defines the procedures to reset the Generator Tracking and Trending data and filament aging calibration to factory values.

Procedure

  1. Select Service Desktop.
  2. Select Hardware ResetScan.
    note:

    This step guarantees Jedi is in Applications mode and Reset TnT will work properly.

  3. Select DIAGNOSTICS from Service Desktop.
  4. Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
  5. Click RESET TNT DATA
  6. Select (highlight) TUBE TNT DATA and GENERATOR TNT DATA .
  7. Click RESET.
  8. Click OK in Attention window.
  9. Click CANCEL in Reset TnT Data window.
    note:

    If process fails, take appropriate corrective action and repeat procedure. If the process fails persistently, perform Initialize Generator Database.

  10. Click DISMISS in Generator Tool - JEDI window.

Finalization

  1. Filament Calibration
  2. Save System State.