Initialize Generator Database

Prerequisites

This document defines the procedures to reset the generator database to factory values.

Procedure

    Notice Image
    • notice
  1. Select DIAGNOSTICS from Service Desktop.
  2. Select GENERATOR TOOL – JEDI, or X-RAY GENERATION, GENERATOR TOOL – JEDI.
  3. Click INITIALIZE DATABASE.
  4. Click OK in Attention window.
  5. Click OK in Initialize Database window.

Finalization

  1. Restore Generator Runtime Parameters (See Save Restore Generator Runtime Parameters)
    note:

    If Save Generator Runtime Parameters failed, perform Reset TnT, Filament Calibration.

  2. Click DISMISS in Generator Tool - JEDI window.
  3. Save System State.