- Date: Apr 22, 2019 12:55:53 AM
Quality Assurance Test
Prerequisites
Overview
This document provides the necessary steps to demonstrate System Image Quality.
1 Phantom Setup
Procedure
- Record phantom serial number in the appropriate GE e4879 Installation form, GE e4879 Certified Component Verification form, or GE Data Report Planned Maintenance Schedule form.
- Perform Center Phantom
- Set Internal Landmark.
2 Standard Test
2.1 Acquire Images
Procedure
- Complete Phantom Setup (See Section 4.1).
- Select NEW PATIENT.
- Type the following entries in the listed Patient Information fields:
- Patient ID: Service
- Patient Name: Quality Assurance Test
- Select Service from Protocol Selection, Anatomical Selector.
- Select MANUFACTURING, IMGSER 20CMQA VCT 64.
- Click CONFIRM to scan Series 1, Series Description: QA1 4x5/120kV/575mA/1s prescription.
- Press START on SCIM/GSCB when it flashes.
- Click NEXT SERIES when Series 1 scans are completed.
- Click CONFIRM to scan Series 2, Series Description: QA3 Heli/0.516/120kV/135mA/1s prescription.
- Press MOVE TO SCAN on SCIM/GSCB when it flashes.
- Press START on SCIM/GSCB when it flashes.
- Click END EXAM.
- If PPS Response window is displayed, click COMPLETE.
2.2 Analysis
Procedure
- Select SERVICE desktop and move Service desktop to left monitor.
- Select IMAGE QUALITY from Service desktop.
- Select IMAGE ANALYSIS and move Image Analysis window to left monitor.
- Select scanned Examination Number, Series Number 1 and Image Number 1 from right monitor List/Select browser.
- Click AUTO 1X radio button on Image Analysis window.
- Click AUTO 1X selection button, and select IMGSER 20QA, QA1 4x5/120kv/57mA/1sec from pull down menus.
- Click ACCEPT.note:
At test completion, a Report window is displayed, which includes specification limits and Pass/Fail results.
note:Test Pass/Fail summary is posted at top of Report window.
note:If test fails, repeat test. If test fails persistently, perform appropriate corrective action.
- Click DISMISS in Report window.
- Select scanned Examination Number, Series Number 3 and Image Number 1 from right monitor List/Select browser.
- Click Image Analysis AUTO 1X selection button, and select IMGSER 20QA, QA3 Heli/0.516/120kv/135mA/1sec, AUTO from pull down menus.
- Click ACCEPT.note:
At test completion, a Report window is displayed, which includes specification limits and Pass/Fail results.
note:Test Pass/Fail summary is posted at top of Report window.
note:If test fails, repeat test. If test fails persistently, perform appropriate corrective action.
- Click DISMISS in Report window.
- Click QUIT in Image Analysis window.
3 Finalization
Procedure
- Record Standard Test Pass/Fail summary in the appropriate GE e4879 Installation form, GE e4879 Certified Component Verification form, or GE Data Report Planned Maintenance Schedule form.
- Store the appropriate GE e4879 Installation form, GE e4879 Certified Component Verification form, or GE Data Report Planned Maintenance Schedule form with system history records.
- Store the Standard Test Quality Assurance results with system history records. The results are stored in a text file on the scanner and can be retrieved at /usr/g/service/state/image_analysis.log.