MTCB Replacement
Prerequisites
Overview
Procedure
- Raise the table to maximum height.
- Remove gantry covers as required.
Refer to
warning- Remove power from table by turning off HVDC Enable, Axial Enable, and 120VAC switches on Service Switch Panel inside gantry.
- Remove right Table Base Cover.
- Remove the two screws holding MTCB Cover in place.
- Insert new MTCB, reassemble and power on Table.
Finalization
- Ensure that proper torque specifications (see Torque Wrench Information) are followed for all fasteners.
- Power up the Table from the Service Switch Panel.
- Perform Flash Download, if prompted to do so.
- Perform Table Characterization Procedures.