Removes the rear endbell from the magnet enclosure.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 30 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Required conditions |
|---|
| Rear cable cover is removed. |
| Rear bridge inner cover is removed. |
| ISC LOTO is applied. |
Procedure
- Disconnect the cooling air hoses from the body coil.
Figure 1. Rear endbell connections
| 1 | Patient cooling air connection |
| 2 | HEX temperature sensor connector |
| 3 | Body coil cooling air connections |
- Disconnect the HEX temperature sensor connector.
| Notice |
|---|
 | Component damage The heat exchanger temperature sensor cable is highly susceptible to damage. Use caution when connecting, disconnecting, or working around the heat exchanger temperature sensor cable. |
- Disconnect the cable connectors from the end of travel switch.
Figure 2. Rear bridge end of travel switch connector
- Disconnect the patient cooling air hose.
- Disconnect the cable connector from the speaker.
- Remove six screws securing the rear endbell to the magnet.
Figure 3. Rear endbell screws
- Remove the rear endbell from the magnet.