• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30FD5C40GYZ
  • Topic ID: id_2019288 Version: 2.5
  • Date: Jul 7, 2021 3:51:55 PM

Removing the left-hand table side cover

Removes the left-hand cover from the dockable table.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
The right-hand table side cover is removed.

Procedure

  1. Remove the three screws securing the side cover to the table.
    Figure 1. Left-hand side cover side screws
  2. Remove the three screws securing the side cover to the table.
    Figure 2. Side cover bottom screws
  3. Remove the left-hand table side cover from the table.
    Note: The acrylic tube is supported by the left- and right-hand table side covers. Make sure the tube is still supported and does not drop when the cover is removed.
    Figure 3. Left-hand side table cover