• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30ED1110GYZ
  • Topic ID: id_2003418 Version: 3.3
  • Date: Jul 7, 2021 3:51:59 PM

Removing the magnet left-hand lower front cover

Removes the lower cover from the left-hand front corner of the magnet.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-50 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

Do the following tasks to complete removing the magnet left-hand lower front cover:
  • Removing the magnet left- and right-hand lower access panels
    Note: Only the left-hand lower access panel must be removed.
  • Removing the magnet service access trim
  • Removing the magnet front service access panel
  • Removing the magnet rear service access panel
  • Removing the magnet service access vent
  • Removing the magnet left-hand lower front cover