Removes the front bridge cover from the front endbell. This task can be performed with the front endbell installed on the magnet or not.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 5 minutes | - |
| Tools and test equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| Nonmagnetic Titanium Service Tool Kit, Small Set | 1 | 5113258 | - |
| Required conditions |
|---|
| Table interface cover is removed. |
Procedure
- Lower the patient table to the down-limit.
- Remove two screws securing bridge cover to front endbell.
Figure 1. Front bridge screws
- Slide the bridge cover downward until the tabs disengage from their tracks on the front endbell.
- Remove the bridge cover.