Removing the bridge cover

Removes the front bridge cover from the front endbell. This task can be performed with the front endbell installed on the magnet or not.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-5 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
Nonmagnetic Titanium Service Tool Kit, Small Set15113258-
Required conditions
Table interface cover is removed.

Procedure

  1. Lower the patient table to the down-limit.
  2. Remove two screws securing bridge cover to front endbell.
    Figure 1. Front bridge screws
  3. Slide the bridge cover downward until the tabs disengage from their tracks on the front endbell.
  4. Remove the bridge cover.