This procedure describes Laser Light check and alignment. Axial
laser light position is at 8mm away from Bridge end. And, cradle is
used for adjustment jig.
Procedure
Set the cradle as follows to use the cradle as alignment jig.
Note: Laser light needs to be adjusted 8mm away from the Bridge end. In this step, joint of cradle end cover will be used as alignment line for laser adjustment.
Move the cradle in so that the “Bridge edge” and
“ “joint of cradle end cover” is aligned.
Press [Landmark].
Move the cradle out until 8 mm is shown.
Note:
S and I shown
on the display changes according to the patient orientation (head
first or feet first).
Note: The right side panel is a mirror image of the left side. Out button is located at the far side from the Magnet bore.
Press [Alignment] button on the control
panel to turn on the laser light.
Figure 1. Cradle preparation
Check for proper alignment.
The axial light should align with
the “joint of cradle”.
The sagittal light should align with
the middle line of the cradle.
If alignment is correct, you are done with this procedure. Otherwise,
continue with the steps that follow.
Note: “Joints of cradle end cover and cradle on each side” are now located at 8mm away from Bridge end.
Figure 2. Proper Alignment
For most of cover, Remove the two snap-in trim ring side sections.
Figure 3. Trim Ring Side Sections
For most of cover, Remove the laser light cover by removing two screws. Cover is held by two poppers and snaps off.
Figure 4. Remove Laser Cover
Change the positioning of the alignment lights by moving the
joystick. Adjust the alignment via the joystick so that the axial
light (x-axis) is aligned to the bridge and endcap joints on both
sides of the bridge and the sagittal light (z-axis) is centered down
the middle of the bridge. See Figure 2 for proper
alignment.
Note: There is only one joystick to adjust the laser light.
Figure 5. Adjusting Laser Lights With Single Joystick