DPP modules RRx1 (P2), RRx2 (P1 (16-CH) and P4 (16-CH)), and RRx3 (PA port E).
Procedure
- P2 DPP RRx1:
- Attach cable C5 (25 pin sub-D, male-female) to J3 at RRx and to J2 of the MCRv tool.
- Attach cable C7 (16w16, male-male) to J2 at RRx and to J4 of the MCRv tool.
- Attach cable C7 (16w16, male-male) to J1 at RRx and to J3 of the MCRv tool.
- P1 DPP RRx2:
- Attach cable C5 (25 pin sub-D, male-female) to J3 at RRx and to J2 of the MCRv tool.
- Attach cable C7 (16w16, male-male) to J2 at RRx and to J4 of the MCRv tool.
- P4 DPP RRx2:
- Attach cable C5 (25 pin sub-D, male-female) to J3 at RRx and to J2 of the MCRv tool.
- Attach cable C7 (16w16, male-male) to J1 at RRx and to J3 of the MCRv tool.
- Attach cable C7 (16w16, male-male) to J1 at RRx and to J3 of the MCRv tool.
- Attach cable C7 (16w16, male-male) to J2 at RRx and to J4 of the MCRv tool.
- Attach cable C5 (25 pin sub-D, male-female) to J3 at RRx and to J2 of the MCRv tool.
- Run the test. See Running the coil datapath diagnostic.
Result
If a test fails:
- Make sure that the connectors are firmly seated.
- Make sure that the screws are properly tightened.
- Rerun the test.
If the test fails again:
- Reseat the connectors.
- Reseat all test cable connections to the MCRv tool connectors.
- Rerun the test.