• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30F50110GYZ
  • Topic ID: id_2003305 Version: 3.1
  • Date: Oct 22, 2021 4:09:57 PM

Doing an ISC coolant warm up scan

Do this before starting RF Amp or UPM calibration if the ICC/CCU temperature is below 29°C. Repeat scan as needed.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-15 minutes-
Tools and test equipment
ItemQuantityPart numberManufacturer
DQA III Phantom12321556-
DQA Phantom Positioner15554497-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.

About this task

It is important to use a phantom with a known geometry, such as the DQA phantom.
Note: Repeat the scan as needed until the temperature is within the required range.

Procedure

  1. Follow the steps below to place the DQA III phantom on the table.
    Figure 1. Phantom positioning on table
    Note: This image is a representative example. Actual systems may vary.
    1. Place the phantom positioner onto the hollow for HNU. R/L direction will be fit to the hollow.
    2. Push the phantom positioner toward the magnet until the two bottom bars reach the end of the hollow.
    3. Place the phantom onto the positioner and verify it is level and not rotated.
  2. Landmark on the center line of the phantom. The laser must be in the middle of the line. Press Advance to Scan.
  3. Click the New Worklist icon.
    Figure 2. New worklist icon
  4. Select geservice for Patient ID.
    Figure 3. Entering a new worklist
  5. Enter for weight: 111.
  6. Click Show All Protocols....
  7. When the Protocol Selector menu opens, click Protocol Details.
  8. Select the Protocol Library: GE.
    Figure 4. Setting protocols
  9. Select Adult.
  10. Click the Template tab.
  11. Select 3 - Plane 2D Localizer and click the arrow to move the selection to the box on the right.
    Figure 5. Protocol selected
  12. Click Accept.
  13. Click Save.
  14. Click Start Exam.
  15. Enter the following image parameters.
    Figure 6. Image parameters
    OptionDescription
    Frequency field of view (FOV) (Freq. FOV)30.0
    Phase FOV1.00
    Slice Thickness5.0
  16. Select Body Coil.
    Figure 7. Coil selection
  17. Click Task and select Change/Select Anatomy.
  18. Select Anatomical Region (Any).
  19. Click Accept.
    Figure 8. Anatomical region
  20. Click Save RX.
  21. Click Scan.
  22. Repeat the scan as needed until the temperature is within the required range.