• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30089A50GYZ
  • Topic ID: id_2026733 Version: 2.1
  • Date: Jul 22, 2021 10:01:32 PM

Configuring DICOM query/retrieve from a network or archive node

Configure DICOM query/retrieve from a network or archive node.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-10 minutes-

About this task

DICOM query/retrieve can be configured by checking the relevant checkbox in SCU Settings and SCP Settings:

Procedure

  1. In the SCU Settings area, select Query/Retrieve. SCU settings apply to all network nodes.
  2. In the SCP Settings area, select Allow Query or Allow to Retrieve. SCP settings apply to network nodes that behave as service class providers. For example, archive node.
  3. Click Save.
  4. Click OK.