Checking the Emergency Stop and Off functions

Follow these procedures to check the Emergency Stop function of the for the Gradient subsystem and RF subsystem, and to check the Emergency Off function of the power for the entire system.

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-45 minutes-
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.