Testing the electrical connections on the ECG cables
Prerequisites
Personnel requirements
Required persons
Preliminary requirements
Procedure
Finalization
1
-
5 minutes
-
Tools and Test Equipment
Description
Quantity
Part Number
Manufacture
Ohmmeter
1
-
-
Procedure
If not already done, disconnect the ECG cables from the PAC and from each other.
Figure 1. PIN Diagram of ECG Lead
Check the impedance of the ECG cables from each of the leads at point A to the other end of that lead at point B. The impedance should be 55k ohms to 63k ohms tolerance.
Check the impedance of the ECG cables from each lead at point to its respective connection at point C. The impedance should be 0 ohms.
Figure 2. Connections A and B on ECG Cables
Connect the white high impedance MRI ECG lead wires to the grey ECG patient cable by matching the colors on the white cable to the colored dots on the grey cable.
Figure 3. PIN diagram of ECG lead
Table 1. PIN diagram of ECG lead
Callout
Color
1
Blank
2
Brown
3
Green
4
Black
5
White
6
Red
Differential pair combinations are
Red (left leg) - White (right arm)
Black (left arm) - Green (right leg)
Plug the open end of the grey ECG patient cable into the ECG port.