• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WHA30A772GYZ
  • Topic ID: id_2001624 Version: 6.6
  • Date: Jul 13, 2021 4:05:26 PM

Calibrating B0 drift

Prerequisites

Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1-4 to 12 hours-
Tools and test equipment
ItemQuantityPart numberManufacturer
Body TLT Loader for 3.0T12360037-
Body TLT Sphere for 3.0T (Pink Solution)12360025-
Service Foam Positioner15554839-
Required conditions
All other applicable calibrations must be complete.
System must be idle for a minimum of 2 hours before functional check.
Gradient Cooling Unit in ICC must be in operation while system is in idle for a minimum of 2 hours before functional check.
Safety

Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:

  • Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
  • Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
  • Comply with all site-specific training and workplace safety requirements.

If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.