CAMERA CALIBRATION
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | - | - | - |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| - | - | - | - |
About this task
This procedure describes the steps necessary to verify and set the camera parameters. Once the display is re-calibrated, it is essential to re-calibrate the camera before the system is used for filming. Although a qualified GE Service Engineer could perform the steps below, it is recommended that the following procedure be performed with the on-site assistance of the camera vendor field engineer.
DASM Interpolation Setup (If DASM is present)
Procedure
- Select the Install soft key from the Service Desktop. When prompted, login using the password operator.
- Select the DASM folder from the top of the Guided Install GUI once it appears.
- Set the DASM Interpolation method to linear.
- Before exiting the GUI, insert the SaveInfo MOD or new MOD into the drive. Go to the top left corner of the Install GUI and select File and then Save GI Configuration to MOD. This process does not create a SaveInfo disk. It just creates a copy of the information already entered in the GUI tabs for use in the next software install.
- Exit the install GUI by selecting File then Quit. If any error conditions still exist, you will be warned that no changes will be made before exiting.
Camera Imaging Look-Up Table
Procedure
- Verify with the camera vendor field engineer that the currently installed camera lookup table is designed to provide perceivably linear contrast for the light box conditions in the customer’s viewing area.
- If not, request that the camera vendor field engineer replace it with the appropriate look-up table.
Camera Maximum Optical Density
Procedure
- Note that for optimal reviewing, the light box luminance of the diagnostic region of the film should be in the range of 50 to 500 nits.
- Vary the maximum optical density setting of the camera to compensate for the light box and to meet this value. A good starting position is a maximum density of 2.8.
Camera Contrast
Procedure
Anatomical Filming
About this task
This portion of the procedure requires the technologist to verify the camera settings with true anatomical images.
Procedure
- Film representative anatomical images to confirm the settings. The image set should include T1 and T2 head images, joint images and c-spines.
- Observe the accuracy of the low-tones, mid-tones and high-tones. If a filmed image is found to not be equivalent then re-calibrate the camera based on the customer’s evaluation.
Finalization
Finalization
No finalization steps.