• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA305F0E20GYZ
  • Topic ID: id_13105970 Version: 3.0
  • Date: Aug 29, 2019 1:39:41 AM

Bird 5000-EX setup for RF Body Calibration

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not ApplicablePart of parent procedure minutesNot Applicable
Topic ID: id_SL13643030-1274795

Setting up RF Coupler for Body Calibration

Procedure

  1. Insert the 50 kW element into the Forward sensor port (blue 4300A374-3). Lock the element in place using the locking tab on the sensor.

    The Reflected port is not used during this procedure. Placing the 5000 W element (blue 4300A374-6) into the Reflected port is just a safety measure, and is not used during body calibration.

    Make sure the arrows on the elements point in opposite directions.

    Figure 1. RF Coupler Setup for Body Mode
  2. Use the DVM to measure the resistance of input to the dummy load. Confirm that the resistance measures 50 ohms +/- 5 ohms.
    Note:

    If the dummy load is out of spec, return the RF kit for repair. Do NOT proceed with the RF amplifier output calibration.

Topic ID: id_SL13643031-1274795

Wattmeter and Dummy Load Setup for Body Channel 1

Procedure

  1. Verify that the scanner is idle. Confirm that the RF ENB switch on the DTX1 exciter is in the Disable (down) position.
  2. Setup the hardware per following steps
    1. Disconnect both channel 1 and 2 body output cables (at J4 and J22) from the front panel of the RF amplifier and set both aside.
    2. Remove cable J21 (RF input for body channel 2) from the body channel input BNC on the front panel of the RF amplifier.
    3. Attach the 3/16-Wavelength Cable Stub (part of the RF Power Measurement Kit) to body channel output J22 to cover exposed RF port that is not being tested.
      Note:

      The 3/16-Wavelength Cable Stub is used as a safety measure to cover and prevent contact with an exposed RF connector. It is not required for successful calibration.

    4. Connect the input of the coupler to RF amplifier J4.
    Note:

    This arrangement will remain in place during UPM calibration and functional check for channel 1 ONLY.

    Figure 2. Hardware setup at RF Amp
  3. Connect the output of the RF coupler to the dummy load.
    Figure 3. Complete Body Output Test Setup for Dual Drive Amplifier
  4. Plug in the wattmeter to use wall power (batteries are often not charged). Connect one end of the wattmeter cable to the sensor connector (RS232 is unused). Connect the other end to the coupler.
  5. Plug the fan for the dummy load into a service outlet located in the lower right part of the ISC. Never use this dummy load without its fan running.
  6. Set the RF ENB switch to ENABLE (up).
Topic ID: id_SL13643032-1274795

Setting up Bird 5000-EX Wattmeter to Measure Forward Body Maximum Power

About this task

This procedure is for the Bird 5000-EX.
Figure 4. Bird 5000-EX Wattmeter

Procedure

  1. Press the ON button on the wattmeter. (If the coupler and sensor are not correctly attached to the amplifier you will see a “No Sensor” indication.)
  2. Select the arrow ▲ button under Scale.
  3. Enter 50000 to set the range.
  4. Select the ENTER button.
  5. Check the setting by selecting the arrow button under Scale.
  6. Press the ENTER button to exit.
  7. Select the arrow button under Fwd Units. Make sure the scale reads W (or kW“on some wattmeters). If the scale reads dBm, select the arrow button under Fwd Units again to toggle the scale to read W (Watts) or kW (kiloWatts) on the TOP, forward power scale. Ignore the BOTTOM, reflected power scale value. It is not used during maximum power calibration.
  8. Select the SHIFT button.
  9. Select the arrow button under Setup.

    This displays the sensor type configured in the wattmeter. This procedure uses sensor type 43, which is optimized for peak power measurement.

  10. Verify that the output reads 43, and then press the ENTER button.
    Figure 5. Wattmeter Display Set to Read
  11. Press the ENTER button.
  12. Press the arrow ▲ button under Type and then toggle to change the unit of measurement to Peak.
  13. All kits require an offset to ensure accurate measurement. Locate the Element Offset Factors card in the kit.

    Find the offset for “3T Dual Drive Body Output”.

    Figure 6. Example Element Offset Factors Card
  14. Press the Offset button on the wattmeter. Enter the numerical value from the offset card and press the Enter button.
    Note:

    To enter a negative value, first enter the numerical value, then the minus symbol (–).

    Figure 7. Wattmeter Set to Display Peak Power
Topic ID: id_14935531

What to do next

Finalization

No finalization steps.