• SIGNA™ Hero 3.0T Service Methods
  • 5852800-8EN Revision 1.0
  • Object ID: 00000018WIA30101E20GYZ
  • Topic ID: id_13106445 Version: 3.0
  • Date: Aug 29, 2019 1:39:45 AM

Agilent Wattmeter and Power Adjustment Kit for RF Body Calibration

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not ApplicablePart of parent procedure minutesNot Applicable
Topic ID: id_SL13643076-1274797

Agilent Wattmeter Setup

About this task

The Agilent wattmeter must be powered on for at least 30 minutes for stable and accurate readings. Plug in the wattmeter and turn it on at the beginning of this procedure.

Procedure

  1. Set up the wattmeter for calibration as shown in the following illustration.
    1. Connect the Agilent sensor probe cable to the sensor.
    2. Connect the sensor to the power reference (PWR REF) port on the wattmeter.
    3. Connect the other end of the sensor probe cable to the Channel A port on the wattmeter.
      Figure 1. Agilent Wattmeter Calibration Setup
      1Front of wattmeter 4Zero/Cal button
      2Power reference (PWR REF) port 5Channel A port
      3Agilent RF power sensor 6Agilent sensor probe cable
  2. On the front of the wattmeter, press Zero/Cal.
  3. Press the Zero + Cal softkey to start the calibration process.

    The wattmeter displays a wait screen while it zeros. While the wattmeter calibrates, a green LED next to the power reference port turns on (indicating RF power is being transmitted from the reference port). When this LED turns off, the calibration is complete.

    If there is a calibration error, an error message appears on the wattmeter screen. If this happens, check the connections to the meter and sensor and press the Zero + Cal softkey to re-run the calibration.

    Note:

    Expect to see a fluctuating low power reading on the wattmeter on the order of ±3.0 dBm (± 2.0 mW) after it is calibrated.

Topic ID: id_SL13643077-1274797

Wattmeter and Dummy Load Setup for Body Mode

Procedure

  1. Verify that the scanner is idle. Confirm that the RF ENB switch on the DTX1 exciter is in the Disable (down) position.
  2. Setup the hardware per following steps.
    1. Disconnect both channel 1 and 2 body output cables (at J4 and J22) from the front panel of the RF amplifier and set both aside.
    2. Disconnect the cable J21 (RF input for body channel 2) from the body channel input BNC on the front panel of the RF amplifier.
    3. Attach the 3/16-Wavelength Cable Stub (part of the RF Power Measurement Kit) to body channel output J22 to cover exposed RF port that is not being tested.
      Note:

      The 3/16-Wavelength Cable Stub is used as a safety measure to cover and prevent contact with an exposed RF connector. It is not required for successful calibration.

    4. Connect the input of the coupler to RF amplifier J4.
    Note:

    This arrangement will remain in place during UPM calibration and functional check for channel 1 ONLY.

    Figure 2. Hardware setup at RF Amp
Topic ID: id_SL13643078-1274797

Setting up Agilent Wattmeter to Measure Forward Body Maximum Power

Procedure

  1. Use a digital multimeter to measure the resistance between the dummy load RF connector center pin and the shield. Confirm a reading of 50 ohms (+/- 5 ohms). Replace the dummy load if the resistance is out of range.
  2. Connect the remaining Agilent power components as shown below.
    Figure 3. Agilent Power Meter Setup for RF Body Calibration
    1RF amplifier (front) 6RF cable to dummy load
    2RF output port 7RF coupler
    3Wattmeter 8Agilent 50 ohm terminator connected to the Reflected port on the coupler
    4Agilent sensor probe cable 9Dummy load with internal fans
    5RF sensor connected to the Incident port on the coupler
  3. Connect one end of the RF sensor to the wattmeter. Connect the other end to the coupler.
  4. Plug the fan for the dummy load into a service outlet located in the lower right part of the ISC. Never use this dummy load without its fan running.
  5. When the power meter has started, press Preset/Local.
  6. Select Default, and then press Confirm.
  7. Press System.
  8. Press Save/Recall.
  9. The meter displays a list of predefined applications.

    (For dual drive RF amplifiers) The meter displays a list of predefined applications. Use the up and down cursor keys to highlight 3T_Dual_Body. Use this setting for both body channel 1 and body channel 2.

  10. Select Recall.
  11. Press Confirm.
Topic ID: id_14934616

Finalization

Finalization

No finalization steps.