- SIGNA™ Hero 3.0T Service Methods
- 5852800-8EN Revision 1.0
- Object ID: 00000018WIA30101E20GYZ
- Topic ID: id_13106445 Version: 3.0
- Date: Aug 29, 2019 1:39:45 AM
Agilent Wattmeter and Power Adjustment Kit for RF Body Calibration
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | Part of parent procedure minutes | Not Applicable |
Agilent Wattmeter Setup
About this task
The Agilent wattmeter must be powered on for at least 30 minutes for stable and accurate readings. Plug in the wattmeter and turn it on at the beginning of this procedure.
Procedure
CAUTION Possible equipment damage!Physical shocks to the sensor and coupler (for instance, dropping them) can damage them or cause them to go out of calibration.Take care when handling the coupler and adapter. When connecting to the sensor, hold the sensor steady and turn the knurled rings on the connectors to tighten them. Take care not to cross-thread the sensor connections.
Set up the wattmeter for calibration as shown in the following illustration.Notice Make sure the serial numbers for the RF coupler and RF sensor match each other. They are calibrated together, and should not be swapped between power measurement kits.- Connect the Agilent sensor probe cable to the sensor.
- Connect the sensor to the power reference (PWR REF) port on the wattmeter.
- Connect the other end of the sensor probe cable to the Channel A port on the wattmeter.
Figure 1. Agilent Wattmeter Calibration Setup 
1 Front of wattmeter 4 Zero/Cal button 2 Power reference (PWR REF) port 5 Channel A port 3 Agilent RF power sensor 6 Agilent sensor probe cable
- On the front of the wattmeter, press Zero/Cal.
- Press the Zero + Cal softkey to start
the calibration process.
The wattmeter displays a wait screen while it zeros. While the wattmeter calibrates, a green LED next to the power reference port turns on (indicating RF power is being transmitted from the reference port). When this LED turns off, the calibration is complete.
If there is a calibration error, an error message appears on the wattmeter screen. If this happens, check the connections to the meter and sensor and press the Zero + Cal softkey to re-run the calibration.
Note:Expect to see a fluctuating low power reading on the wattmeter on the order of ±3.0 dBm (± 2.0 mW) after it is calibrated.
Wattmeter and Dummy Load Setup for Body Mode
Procedure
DANGER FATAL ELECTRIC SHOCK HAZARD.THE RF AMPLIFIER IS NOT DISABLED.TO PREVENT FATAL ELECTRIC SHOCK, ENSURE THE RF AMPLIFIER IS DISABLED WHILE CONFIGURING THE CALIBRATION TOOLS.CAUTION
Electrical/RF burn hazardRF burns can occur as a result of disconnecting heliax cables while the system is scanning.To prevent possible RF burns when disconnecting heliax cables from the RF amplifier, make sure that the system is not manually prescanning or scanning. Make sure that the scan desktop icon displays the “idle” message and the RF enable switch is in the “off” position.
Verify that the scanner is idle. Confirm that the RF ENB switch on the DTX1 exciter is in the Disable (down) position.Notice Possible equipment damage. Do not connect/disconnect the wattmeter sensor while the meter is ON. Always turn the meter OFF before plugging/unplugging the sensor.
Setup the hardware per following steps.CAUTION Equipment Damage!If the body coil is connect to the RF amplifier and the wrong channel is selected, damage to the DTRSW can occur.Disconnect the unused body channel during testing and calibration of the RF amplifier.- Disconnect both channel 1 and 2 body output cables (at J4 and J22) from the front panel of the RF amplifier and set both aside.
- Disconnect the cable J21 (RF input for body channel 2) from the body channel input BNC on the front panel of the RF amplifier.
- Attach the 3/16-Wavelength Cable Stub (part of the RF Power
Measurement Kit) to body channel output J22 to cover exposed RF port that is not being tested. Note:
The 3/16-Wavelength Cable Stub is used as a safety measure to cover and prevent contact with an exposed RF connector. It is not required for successful calibration.
- Connect the input of the coupler to RF amplifier J4.
Note:This arrangement will remain in place during UPM calibration and functional check for channel 1 ONLY.
Figure 2. Hardware setup at RF Amp 
Setting up Agilent Wattmeter to Measure Forward Body Maximum Power
Procedure
Use a digital multimeter to measure the resistance between the dummy load RF connector center pin and the shield. Confirm a reading of 50 ohms (+/- 5 ohms). Replace the dummy load if the resistance is out of range.Warning Electric shock risk!pay close attention to the connections to the dummy load. Failure to connect the dummy load correctly could result in electric shock.- Connect the remaining Agilent power components as shown below.
Figure 3. Agilent Power Meter Setup for RF Body Calibration 
1 RF amplifier (front) 6 RF cable to dummy load 2 RF output port 7 RF coupler 3 Wattmeter 8 Agilent 50 ohm terminator connected to the Reflected port on the coupler 4 Agilent sensor probe cable 9 Dummy load with internal fans 5 RF sensor connected to the Incident port on the coupler - Connect one end of the RF sensor to the wattmeter. Connect the other end to the coupler.
Plug the fan for the dummy load into a service outlet located in the lower right part of the ISC. Never use this dummy load without its fan running.Notice Do not apply RF power to the dummy load without its fan in operation. Doing so will result in catastrophic damage to the dummy load due to overheating.CAUTION Possible equipment damage!Physical shocks to the sensor and coupler (for instance, dropping them) can damage them or cause them to go out of calibration.Take care when handling the coupler and adapter.
When the power meter has started, press Preset/Local.Notice Make sure the serial numbers for the RF coupler and RF sensor match each other. They are calibrated together, and should not be swapped between power measurement kits.- Select Default, and then press Confirm.
- Press System.
- Press Save/Recall.
- The meter displays a list of predefined applications.
(For dual drive RF amplifiers) The meter displays a list of predefined applications. Use the up and down cursor keys to highlight 3T_Dual_Body. Use this setting for both body channel 1 and body channel 2.
- Select Recall.
- Press Confirm.
Finalization
Finalization
No finalization steps.