This is one of the procedures required to install the host system software. Follow this procedure after installing Linux Host OS software.
Prerequisites
| Personnel requirements |
|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | - | 30 minutes | - |
| Tools and Test Equipment |
|---|
| Item | Quantity | Part number | Manufacturer |
|---|
| See the FRU manual for part numbers. |
| MR Applications Software DVD | 1 | - | - |
| Required conditions |
|---|
| Linux OS has been installed. |
Procedure
- After the system reboots, the Welcome to MRHost screen displays. Log in using these parameters:
- User ID: root
- Password: operator
- Insert the MR Applications Software DVD into the appropriate DVD drive when prompted. Click OK after the DVD drive LED stops blinking.
- From the Select your locale window, select the correct language for the site. Click OK.
A series of messages and a progress bar report the progress of the installation. When complete, the Guided Install screen appears.
- Choose one of the following after the prompt Do you want to load GI Configuration from DVD? appears.
- For new system installations, close the drive and select No. Use the information recorded in System and hardware configuration to complete this section.
- For reinstallation, insert the Save Info media and select Yes after the DVD drive LED stops blinking. Load GI restores Guided Install information. When the process is complete, reinsert the MR applications software DVD and continue to HIPAA/Security.
Note: If upgrading to a new configuration, check to make sure that the same configuration is applied in Guided Install. See
System and hardware configuration for more information.
- Select security level.
- Go to .
- Select the security option appropriate for the site per the customer.
Most installations will use typical security settings. The Highest Security Settings requires the SIGNA Secure Advanced option key.
Follow these steps to select the
SIGNA Secure Advanced security level:
- Continue the software installation with the Typical Security Settings until applications software load completes.
- Install the SIGNA Secure Advanced option key and complete the remaining software installation tasks.
- Relaunch Guided Install to update the security setting.
- Reboot the system
Note: SIGNA Secure Advanced disables InSite connectivity.
Note: Once the highest security level is selected, it cannot be changed without performing a software reload.
Refer to Security levels for more information.
- Proceed to the Verification tab. Review the entries on the page to determine if all entries passed verification (the Install Software button will not be active if there are any failures). Failed entries result from missing or improperly formatted entries.
- Go to the tab with the error and correct it. Click Install to open the Verification tab again.
- If all entries pass verification, click Install Software to begin the MR applications load. This process takes 15-20 minutes.
Note: A prompt appears reminding you to configure the NTP server (if present) through the Network Info tab. It also reminds you to verify that the current time and time zone are configured properly on the Set Time/Date tab.
If both are correct, select Confirm. If not, select Abort and configure the NTP server (if present) and the current time and time zone.
- When the Service License screen appears, leave it as-is.
The Service License screen is displayed continuously when the user logs on as root.
- The Save/Restore screen appears once the MR Applications load is complete.
- Remove the MRApps DVD from the drive. Insert the SaveInfo DVD and restore information if applicable.