Setting up 3.0T GEM Peripheral Vascular Array (PVA)
Use this procedure to set up the left, right, and upper GEM Peripheral Vascular Array (PVA) for the Multi-Coil Quality Assurance (MCQA) check.
Prerequisites
Note: Coils do not ship with phantoms. Phantoms come in a unified phantom set with the system.
| Personnel requirements | |||
|---|---|---|---|
| Required persons | Preliminary requirements | Procedure | Finalization |
| 1 | - | 90 minutes | - |
| Tools and test equipment | |||
|---|---|---|---|
| Item | Quantity | Part number | Manufacturer |
| TL Unified Phantom (for 3.0T) | 2 | 5343347-2 | - |
| Large Cylindrical Unified Phantom (for 3.0T) | 1 | 5342679-2 | - |
| Small Cylindrical Unified Phantom (for 1.5T and 3.0T) | 1 | 5342680 | - |
| PA Phantom Positioner for Both 1.5T and 3.0T | 1 | 5405748 | - |
| PV Phantom Positioner for Both 1.5T and 3.0T | 1 | 5405749 | - |
| Safety |
|---|
|
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed. |
About this task
- Do the MCQA test for the 3.0T GEM PVA upper configuration.
- Do the MCQA test for the 3.0T GEM PVA left configuration.
- Do the MCQA test for the 3.0T GEM PVA right configuration.