About this task
Important: When secure DICOM push is configured in one-way TLS/SSL authentication, DICOM ping will only succeed from MR to the networked node and will fail from the networked node to MR. This is expected because the networked node does not have a trusted certificate from the MR scanner. DICOM ping will succeed both ways in secure mode only when two-way TLS/SSL authentication is configured.
Procedure
In the Configure Network Hosts window, click Add.
Figure 1. Add a remote host
Configure the network node by choosing one of the following:
Option
Description
DICOM over TCP/IP (Legacy method unsecure)
Enter the host name, display name, IP address, port, and AE title information of the server. The default port for other GE devices is 4006.
Select the Archive Node option and configure the Storage Commitment Host Details.
DICOM over TLS (secure, available with SW version DV29.0 and later only)