Table Top Height Adjustment

Prerequisites

This procedure is applicable to the following table products:

  • Signa Lightweight Patient Transport Table

  • Signa Oncology Patient Transport Table Option

  • Signa OR-Compatible Patient Transport Table Option

Procedure

  1. Dock Patient Transport. If Dock has not yet been mounted to magnet, complete Dock Centering and Dock Hardware Adjustments procedure.
  2. Pump table top to fully raised position and move the cradle completely into the magnet bore to expose the table top height adjusting screw.
  3. Use a small screwdriver to carefully remove cover plug on table top. See Figure 1.

    Figure 1. REMOVING COVER PLUG

  4. Adjust screw in table top to set table top height equal to bridge height.
  5. Replace cover plug.