CRADLE GUIDE RAIL LATCH ADJUSTMENT
Prerequisites
The cradle is locked in home position on the Fixed Table by a latch in the cradle, which engages a notch in the cradle guide strip near the end of the Fixed table. The cradle should be able to just engage this notch, without extra force, when pushed to home position on the table top, yet it must engage it tight enough so the table top (cradle) will not move on the Patient table freely when it is not engaged with LPCA.
1 Cradle Guide Rail Latch Adjustment when Table is docked to Gantry
Procedure
- Press Table up pedal on the Dock FRP cover, take the table to the top most condition. Press the Cradle in button on the Magnet Cover, Move the cradle inside the Bore by 400mm approximately.
- Loosen five screw securing guide rail.
- Move Guide rail up or down as required to create clearance for cradle movement.
- Move Guide rail forward or back word as required for cradle locking.
Figure 1. Adjustment of Delrin Guide

- Press the Cradle out button on the Gantry, bring back the cradle to home position.
- Ensure that the cradle latches to the delrin guide and also
check for the Cradle Play, by moving the Cradle. If adjustment is
not good, readjust the guide following the same steps.
Figure 2. Checking Cradle latch adjustment

2 Cradle Guide Rail Latch Adjustment when Table is not docked to Gantry
Procedure
- Slide out the cradle by 400mm approximately.
- Loosen five screw securing guide rail.
- Move Guide rail up or down as required to create clearance for cradle movement.
- Move Guide rail forward or back word as required for cradle locking.
Figure 3. Adjustment of Delrin Guide

- Take back the cradle to home position. Ensure that the cradle
latches to the delrin guide and also check for the Cradle Play, by
moving the Cradle. If adjustment is not good, readjust the guide following
the same steps.
Figure 4. Checking Cradle latch adjustment

3 Finalization
Finalization
No finalization steps.