Bleeding Patient Transport Hydraulics
Prerequisites
Procedure
- Unlock patient transport and remove from magnet room.
- Raise patient transport table to highest elevation with Up foot pedal.
- Remove patient transport side covers on right side (when facing magnet enclosure).
- Lower the safety latch by disconnecting spring from scissors assembly
(see Figure 1).
Figure 1. LOWERING SAFETY LATCH

- Release hydraulic pressure by gently pressing Down foot pedal until table is supported by safety latch.
- Hold the Down valve on the lift cylinder open while pumping the Up foot
pedal at the same time. This can be done by one person by propping the Down
valve in the open position using a standard blade screwdriver between the
Down valve and the cylinder block. See Figure 2 for location of Down valve.
Figure 2. LOCATION OF HYDRAULIC FILTER

- After all the air bubbles pass though the clear tubes (back to the reservoir), stop pumping and release the Down valve.
- Remove the screwdriver to close the Down valve.
- Purge the lift cylinder by pumping the table up to the top.
- Open the bleed screw at the cylinder top a few turns until oil flows steadily from the screw after the air bubbles purge.
- Tighten bleed screw (just snug).
- With the table at highest elevation, raise the safety latch and reconnect spring to scissors assembly.
- Lower table to its lowest position using Down foot pedal.
- Fill table reservoir with hydraulic oil (46-230006P2)
so that fluid is 1/4 to 1/2 inch above full line when table is at its lowest
position (see Figure 3).
Figure 3. PATIENT TABLE RESERVOIR FILL LEVEL

- Replace side covers.
Finalization
No finalization steps.