PSE Board Level Diagnostic

1 Disclaimer

Depending on your service agreement, not all service tools, diagnostics, and utilities referenced in this document may be accessible. Contact your sales person for information on available service license packages.

2 Diagnostic Description

2.1 Diagnostic Path

Diagnostics > System Function > Acquisition/Pulse Generation > Data Acquisition

Diagnostics > Hardware Location > System Cabinet > CAM > Data Acquisition

Figure 1. PSE Board Level Diagnostic

2.3 Components Tested

PSE board functionality

2.4 Requirements

No requirements.

2.5 Block Diagram

Figure 2. PSE Block Diagram

To view or print a larger view of this diagram, click 11942524.pdf.

2.6 Test Sequence

  1. Click Calculate Time to get an estimate of the time it will take to run the selected diagnostics.

    • Short Tests: The short test performs a pattern of walking ones and zeros.

    • Long Tests: The long test performs a pattern of walking ones, walking zeroes, 0xaaaa, 0x5555, and possibly others. As a result, selecting the long test results in a more exhaustive test. If the problem is highly intermittent, try a long test.

    note:

    Do not change the Test Options settings from their default settings unless instructed to do so by a GE Healthcare engineer. There are very few circumstances when these settings ever need to be changed.

  2. Click Run to initiate the test.

2.7 Expected Results

Test Results

Diagnostic results display the test name with a status of PASS or FAIL. If you receive a FAIL status, click the error number in the 1st Error column to view error message details.

Test Status

This window informs you of the progress of the test from the initial scan to the completed diagnostic.

2.8 Finalization

Before scanning, perform a TPS reset.