o.37.5 Body Slice Thickness (Coronal, 3mm)

SCAN OPTION INPUT

RX MANAGER . Place cursor in Series list area.

Press right mouse button, then

Copy Series, Paste Series.

Double-click left mouse button

on new series to activate it.

See Note.

IMAGING PARAMETERS

Plane >Coronal

SCANNING RANGE

Spacing 77

ADDITIONAL PARAMETERS Graphic Rx

Position cursor at P= 0 mm

This should coincide with the center of the phantom.

Start

Position cursor 80 mm away (posterior) from list location.

End

See procedure for additional information.

AQUISTION TIMING

Freq 256

Phase 256

NEX 4 (1.0T)

2 (1.5T)

Phase FOV 1.00 (default)

Freq Dir >S/I

Auto Center Freq >Peak

(lowest window) Save Series

RX MANAGER Prepare to Scan

note:

Only values different from previous series are shown. All other values Should be the same as the previous series.