o.37.5 Body Slice Thickness (Coronal, 3mm)
SCAN OPTION INPUT
RX MANAGER . Place cursor in Series list area.
Press right mouse button, then
Copy Series, Paste Series.
Double-click left mouse button
on new series to activate it.
See Note.
IMAGING PARAMETERS
Plane >Coronal
SCANNING RANGE
Spacing 77
ADDITIONAL PARAMETERS Graphic Rx
Position cursor at P= 0 mm
This should coincide with the center of the phantom.
Start
Position cursor 80 mm away (posterior) from list location.
End
See procedure for additional information.
AQUISTION TIMING
Freq 256
Phase 256
NEX 4 (1.0T)
2 (1.5T)
Phase FOV 1.00 (default)
Freq Dir >S/I
Auto Center Freq >Peak
(lowest window) Save Series
RX MANAGER Prepare to Scan
Only values different from previous series are shown. All other values Should be the same as the previous series.