o.37.1 Body Slice Thickness (Sagittal, 3mm)

SCAN OPTION INPUT

PATIENT REGISTER . New Pt

PATIENT INFORMATION

Patient Id geservice

Patient Name body slice thick

Weight (Lb) 111

Landmark

Landmark >Sternal Notch

PATIENT PROTOCOLS Patient Position

PATIENT POSITION

Patient Position >Supine

Patient Entry >Head First

Coil ...BodyAccept

IMAGING PARAMETERS

Plane >Sagittal

Mode >2D

Pulse Seq ...Spin EchoAccept

Imaging Options none

Psd Name no entry

Protocol no entry

SCAN TIMING

# of Echoes 1 (default)

TE 25

TR 300

Bandwidth 7.81 (1.0T) 15.63 (1.5T and 3.0T)

SCANNING RANGE

FOV 20

Slice Thickness 3

Spacing 77

R/L Start R80

R/L End L80

# of Slices 3 (default)

P/A Center 0 (default)

I/S Center 0 (default)

Table Delta 0.00 (default)

ACQUISITION TIMING

Freq 256

Phase 256

NEX 4 (1.0T)

2 (1.5T and 3.0T)

Phase FOV 1.00 (default)

Freq Dir >A/P

Auto Center Freq >Peak

(lowest window) Save Series