o.28.3 Full Field Distortion (Coronal)
SCAN OPTION INPUT
RX MANAGER New Series
PATIENT POSITION
Patient Position >Supine
Patient Entry >Head First
Coil ...BodyAccept
IMAGING PARAMETERS
Plane >Coronal
Mode >2D
Pulse Seq ...Spin EchoAccept
Imaging Options none (default)
Psd Name no entry
Protocol no entry
SCAN TIMING
# of Echoes 1 (default)
TE. 25
TR. 300
Bandwidth 15.63
SCANNING RANGE
FOV 48
Slice Thickness 20
Spacing 1.5
Start See procedure
End See procedure
# of Slices 1 (default)
S/I Center 0 (default)
R/L Center 0 (default)
Table Delta 0.00 (default)
ACQUISITION TIMING
Freq. 256
Phase 256
NEX 1 (1:23)
Phase FOV 1:00 (default)
Freq Dir >S/I
Auto Center Freq. >Peak
(lowest window) Save Series
RX MANAGER Prepare to Scan