Identifying the Image Compute Node (ICN)

1 Disclaimer

Depending on your service agreement, not all service tools, diagnostics, and utilities referenced in this document may be accessible. Contact your sales person for information on available service license packages.

2 Utility description

2.1 Utility path

Utilities > Toolbox > VRE > ICN Identification

2.2 Purpose

This tool applies to systems with multiple ICN configurations only. Each ICN has a host name. This utility identifies each ICN and its host name. This is a helpful identifier when conducting troubleshooting and repair procedures.

2.3 Components tested

No components tested.

2.4 Requirements

System with multiple ICNs.

2.5 Test sequence

  1. Navigate to the ICN Identification utility. Select the identification time from the drop-down menu.

    The default time is 30 seconds. The LED can flash for the maximum length of five minutes.

  2. Click ICN1 or ICN2. The ICN selected is identified by a flashing LED on the front chassis.

2.6 Results

The ICN selected is identified by a flashing LED on the front chassis (including systems with silver Sun ICNs).

For systems with a black Dedicated Computing ICN, only the MAC address displays. Confirm this address is correct with the ICN MAC address label located on the back of the ICN.