The purpose of this diagnostic is to test the disk throughout
for both of the hard disks on an ICN. Hard disks store acquisition
data for certain applications in System. This diagnostic essentially
verifies the ICN hard disk performance is fast enough to keep up with
the incoming data during acquisition.
3 Components Tested
ICN hard disks
4 Requirements
The ICN is turned on and has at least one network connection
from the ICN motherboard Ethernet port to the Host.
5 Block Diagrams
Figure 2. System Cabinet and GOC Diagram
6 Test Sequence
Select the ICN Hard Disk Stress diagnostic from the VRE Hardware Diagnostics screen.
Click Run.
note: Once you click Run, do not click anywhere else in the Common Service Desktop until the test is complete. If you need to abort the test prior to its completion, you must click Stop.
Ensure that the ICN passes the diagnostic by reviewing the status in the Test Results table.
7 Expected Results
This diagnostic has either a passed or failed status. If the
diagnostic has failed, please review the error log and corresponding
extended error messages for subsequent actions. Additional responses
to the diagnostic's failed result are outlined in Table 1.