All media except for Save Info DVD must be removed during the loading of the application software, disk partitioning, and application software load. This includes other DVDs, MODs, and USB devices (for example, service keys). Failure to remove other media can cause software load failure.
System monitor turned on.
All USB devices (including service keys) are removed.
Any non-OS media is removed.
Safety
Before working in any GE Healthcare MR suite or performing any GE Healthcare service procedure, you must:
Have read and understood all hazard conditions and safety requirements in the latest revision of the GE Healthcare MR Service Safety Manual (5452735).
Have successfully completed all relevant GE Healthcare Environmental Health and Safety (EHS) courses (or for non-GE employees, equivalent workplace training courses).
Comply with all site-specific training and workplace safety requirements.
If you have any safety concerns at any time, do not begin work or immediately stop work and move to a safe location. Immediately contact your supervisor or site safety officer for instructions on how to proceed.
Procedure
Insert the USB software device required for this procedure.
Note: A USB-3 port is recommended. The port has a blue-colored plug or is labeled on the box.