Prerequisites
Table 1. Personnel requirements| Required persons | Preliminary requirements | Procedure | Finalization |
|---|
| 1 | Not Applicable | 40 minutes | Not Applicable |
Table 2. Tools and test equipment| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|
| Linux OS Software
DVD | 1 | - | - | - |
Procedure
- Insert the Linux OS Software DVD into the appropriate DVD drive.
- Restart the host computer.
- From the HELiOS based GEHCMR Operating System (MR
OS) Install screen, select 1. GEHCMR – MR Auto
Install and press Enter.
The Welcome to HELiOS for x86_64 screen
displays.
Note:
During the beginning of this process (checking dependencies,
formatting disk drive, etc.), there are two active buttons in the
lower right: ←Back and →Next . Do not select these buttons. Later in the
process, these buttons are greyed out (inactive).
The software load proceeds through some checks and then begins
the OS load. No responses are required.
A progress bar shows the percentage complete. This process takes
about 15 or 20 minutes.
- When the software load is complete, a message appears stating
that installation has completed successfully.
- Eject and remove the DVD.
- Select Reboot.
The OS software load takes about 20 minutes. A progress bar
displays during the reboot sequence.