- Optima MR450w BASE 1.5T System Service Methods
- 5690012-2EN Revision 3
- 00000018WIA30EA6E20GYZ
- id_131061714.2
- Nov 5, 2020 1:31:35 PM
Echo Planar Test (EPT)
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 2 hours | 5 minutes |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| 100 mm Sphere Phantom for 1.5T | 1 | - |
46-317586G1 | - |
| EPI Foam Positioner, Body | 1 | - |
2170481 | - |
| EPI Foam Positioner, Body | 1 | - |
2170481 | - |
| Service Filler Panel | 2 | - |
5344577 | - |
| ||||
About this task
The Echo Planar Test (EPT) combines the functionality of the B0 Dither Calibration and Group Delay Calibration into a single tool with a simple user interface. The output is suitable for remote support access, trending, and automatic checking of results against acceptance limits. EPT automatically updates the Echo Planar Image (EPI) calibration files.