About this task
This procedure provides instructions to calibrate the maximum power for the RF amplifier and also calibrate and do a check of the power monitor calibration for each channel using the UPM Tool.
Use the UPM Tool to process each channel. For example, when you select head mode, the tool executes the RF amplifier calibration and, if the amplifier passes, it proceeds to the UPM calibration and then to the functional check. The process is order-dependent.
For RF amplifier calibration, process the following in order:
- Body channel .
- Head channel.
Buttons change color to keep track of task status:
- Grey: Not started.
- Blue: In progress.
- Red: Error encountered or failed specs.
- Green: Successfully completed and passed.
Click
File for the following menu options:
- View Results: Open previous results file.
- View Config: View RF and power monitor configuration file details.
- Backup Cal Files: Backup upm*Cal.cfg files.
- Restore Cal Files: Restore upm*Cal.cfg files.
Click
Tools for the following menu options:
- Reset Average Power: Run utility to reset power monitor power.
- Check Cal Status: Display status of current power monitor calibrations.
For each channel, adjust the RF amplifier output power first, then calibrate UPM forward and reflected power, and finally run a UPM functional check.