1.5T Neurovascular Array Setup for SNR Test
Prerequisites
| Required persons | Preliminary requirements | Procedure | Finalization |
|---|---|---|---|
| 1 | Not Applicable | 15 minutes | Not Applicable |
| Item | Quantity | Effectivity | Part number | Manufacturer |
|---|---|---|---|---|
| Phantom and positioner kit | 1 | - |
5117092-6 | - |
About this task
Follow this process to prepare for the automated SNR test using the 1.5T NV Array by Medrad (M3087JJ).
Procedure
Finalization
No finalization steps.