• Optima MR450w BASE 1.5T System Service Methods
  • 5690012-2EN Revision 3
  • Object ID: 00000018WIA308BFD20GYZ
  • Topic ID: id_13106852 Version: 3.1
  • Date: Mar 19, 2020 6:17:59 AM

1.5T HD 8-Channel Neurovascular Array Setup for MCQA Test (Invivo)

Prerequisites

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
1Not Applicable25 minutesNot Applicable
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Large Cylindrical Unified Phantom1 Unified Phantom Set

5342679

-
10 cm Spherical Unified Phantom1 Unified Phantom Set

46-317586G1

-
17 cm TLT Spherical Unified Phantom1 Unified Phantom Set

46-265826G6

-
Unified Phantom Positioner1 Unified Phantom Set

5344806

-
Table 3. Required conditions
ConditionReferenceEffectivity

For Discovery MR450 and Optima MR450w, the following coil configuration names must be installed to run this tool: 8NVHEAD_A and 8NVARRAY_A. For HDx, HDxt, and HDe, the following coil configuration names must be installed to run this tool: 8NVHEAD_A and 8NVARRAY_A.

--

About this task

Follow this process to prepare for the automated SNR Test using the 1.5T 8-Channel Neurovascular Array Coil by Invivo (M3335M).

Note:

Coils do not ship with phantoms. Phantoms come in a unified phantom set with the MR system.

Unified Phantom Procedure

Procedure

  1. Place the base of the NV array coil on the patient table and latch it to the table as shown in Figure 1.
    Figure 1. NV Array Coil on Patient Table
  2. Place the phantom positioner in the coil as shown in Figure 2.
    Figure 2. Phantom Positioner in the Coi
  3. Place the phantoms in the positioner as shown in Figure 3.
    Figure 3. Phantoms in the Positioner
  4. Carefully align the connector pins on the base of the coil with the top part of the coil and then latch both the coil sections into place as in Figure 4. (The scanner will not operate if the coil sections are not latched correctly).
    Figure 4. Connector Pins Aligned on Base of Coil
  5. Connect the coil to Port A of the system LPCA. Landmark the coil at the cross mark as shown in Figure 5 and advance to scan.
    Figure 5. Landmark the Coil
  6. Run the Multi-Coil Quality Assurance (MCQA) Tool with the Unified Phantom Set.

Finalization

Finalization

No finalization steps.