• Optima MR450w BASE 1.5T System Service Methods
  • 5690012-2EN Revision 3
  • Object ID: 00000018WIA30131F20GYZ
  • Topic ID: id_13106723 Version: 3.0
  • Date: Aug 29, 2019 1:57:32 AM

HD 8Ch Foot Ankle Coil Troubleshooting Tips

Table 1. Personnel requirements
Required personsPreliminary requirementsProcedureFinalization
10 minutes60 minutes0 minutes
Table 2. Tools and test equipment
ItemQuantityEffectivityPart numberManufacturer
Digital Volt Meter1---
Table 3. Required conditions
ConditionReferenceEffectivity

The following coil configuration names must be installed to run this tool: HD Foot Ankle

--

The following tips can be used to troubleshoot common problems with the 1.5T HD 8Ch Foot Ankle by Invivo, catalog M3340CD.

Receiving No Signal

Problem:

You are unable to pre-scan or are scanning and yet receiving no signal.

Possible Solution:

  • Verify that you have selected the appropriate system coil selection. Refer to the Operators Manual for additional information.

  • Verify the green light above Port A is illuminated. This indicates the coil is properly plugged into the system.

  • Verify that the landmark is correct and that the cradle has not unlatched.

  • Verify that the scan locations and any FOV offsets are correct.

  • Perform a continuity check on the external cable (below). The continuity check must be performed by a GE authorized Service Engineer.

  • Verify that the coil is positioned with the cable exiting towards the bore and that the system and coil polarity match.

  • If you still cannot get a signal, try to scan (transmit and receive) with the body coil. For this test, be sure to remove the imaging coil from the magnet bore before you scan with the body coil. If you still receive no signal, the problem probably lies with your MR system. If the scan completes successfully, there is probably a problem with the Invivo coil. Contact GE for further assistance. If you are unable to scan with the substitute coil, there may be a system problem related to this particular coil type.

Image Quality

Problem:

Image quality is not what you expect it to be given the parameters selected.

Possible Solution:

  • Review the selected protocol. If you are performing the Periodic Quality Assurance, be sure your protocol is identical to the protocol provided in your Systems Recommended Protocols. If you are performing diagnostic images, you may need to increase NEX or FOV.

  • Perform a continuity check on the external cable (below). The continuity check must be performed by a GE authorized Service Engineer.

  • Verify that there are no loops in the cables.

  • Verify that there are no metal or ferromagnetic objects close to the coil, patient or magnet (i.e., safety pin, hair pin).

  • Verify that the coil is properly positioned.

  • Verify that your center frequency is within the frequency adjustment range for your system.

  • Verify that the R1, R2 and TG values from the pre-scan are within normally expected ranges.

  • Cleaning the electrical contacts that connect the upper and lower halves with Isopropyl Alcohol and a cotton swab has been demonstrated to improve the coil’s SNR on occasion. If you have not done so already, perform a MCQA (Multi Coil Quality Assurance) Test. If the values you obtain do not fall within normal operating parameters, investigate this further by performing a phantom scan with the body coil. For this test, be sure to remove the imaging coil from the magnet bore before you scan with the body coil. If you still have the same problems, there is probably an MR system problem. If the body coil scan is satisfactory, acquire a scan using both another coil of the same type plugged into the same system port (port A). If the image quality is visibly improved, there may be a problem with the Invivo coil. Contact GE for further assistance. If the image quality still suffers, there may be a system problem related to imaging with this type of coil.

Artifacts

Problem 1:

There is a black line or signal void on the image.

Possible Solution for problem1:

Verify that there is no metal present in the area being scanned in or on the patient.

Problem2:

Some or all of the images appear shaded or exhibit uneven signal or banding.

Possible Solution for problem2:

Confirm that no metallic objects are located nearby, outside the FOV. This is especially important on images utilizing Fat Saturation. If Fat Saturation is being used, verify that the CFA fine adjustment has been optimized.

External Cable Wear

Problem:

The system will not recognize the coil or scan with the coil attached.

Possible Solution:

Perform the cable continuity check. As follows:

  1. RF Signal Channels Check: The center pins of the System Connector receive coaxial connectors are extremely fragile. Do not attempt to make any measurements at these center pins.
    Note:

    With the coil cable disconnected, use a multimeter to assure an open circuit condition exists between the designated pin and GND for each signal channel shown in Table 1.

    Table 4. Coil Cable Connector Short Circuit and Open Circuit Check
    Signal Channel Coil Cable Connector Pin (Illustration 1) System Connector (Illustration 2)
    RF 1 A17 C1
    RF 2 A15 C2
    RF 3 A10 C3
    RF 4 A21C4
    RF 5 A4 D1
    RF 6 A8 D2
    RF 7 B12 D3
    RF 8 A13D4
    GND

    NOTE: The following pins are common to GND:

    A2-A3, A5, A7, A9, A11-A12, A14, A16, A19-A20, A22-A23,

    B1, B3-B6, B8-B11, B13-B18, B20-B21, and B23-B24.

    K1-K8, M6, MC Shield
    Figure 1. Coil Cable Connector

    If one or more of the readings indicate a short, replace the cable. If all of the above readings indicate an open circuit condition, proceed to the PIN Diode Control Channels Check.

  2. MC bias line (DC line) Check: With the coil cable disconnected, use a multimeter to determine continuity between the designated pins for each signal channel below. Also verify that the control pins are not shorted to GND.
    Table 5. Coil Cable Connector Continuity Check
    Signal ChannelCoil Cable Connector (Illustration 1) System Connector (Illustration 2)
    MC bias 1 A18 J1
    MC bias 2 B22 J2
    MC bias 3 B2 J3
    MC bias 4 B19 J4
    MC bias 5 A6 J5
    MC bias 6 B7 J6
    MC bias 7 A1 J7
    MC bias 8 A24J8
    GND

    NOTE: The following pins are common to GND:

    A2-A3, A5, A7, A9, A11-A12, A14, A16, A19-A20, A22-A23

    B1, B3-B6, B8-B11, B13-B18, B20-B21, and B23-B24.

    K1-K8, M6, MC Shield
    Figure 2. System Connector

    If one or more of the readings are > 3 ohms, replace the coil cable. If all readings are < 3 ohms, proceed to the PIN Diodes Check.

  3. PIN Diodes Check With the coil cable connected to the coil, use a multimeter to measure the diode drop across the coil connector pins as detailed below. Use the System Connector illustration above for pin identification while performing this procedure.
    Table 6. Pin Diode Expected Readings
    System Connector
    Positive LeadNegative LeadReading (V)
    J1GND[MC Shield]1.7 ± 0.2
    GNDG1J1Open
    J2GND[MC Shield]1.7 ± 0.2
    GNDG2J2Open
    J3GND[MC Shield]1.7 ± 0.2
    GNDG3J3Open
    J4GND[MC Shield]1.7 ± 0.2
    GNDG4J4Open
    J5GND[MC Shield]1.7 ± 0.2
    GNDG5J5Open
    J6GND[MC Shield]1.7 ± 0.2
    GNDG6J6Open
    J7GND[MC Shield]1.7 ± 0.2
    GNDG7J7Open
    J8GND[MC Shield]1.7 ± 0.2
    GNDG8J8Open

Mechanical Parts Wear

Inspect the coil for visible signs of wear. If the coil housings will not seat with each other or if the latch mechanism will not work properly, contact Service.

Finalization

No finalization steps.