Updates added for Rev 27, May 2022 release:
-
Added Finalization Step "Geometry Verification" to Gradient Switch Replacement Procedure tacrea3
Updates added for Rev 26, January 2019 release:
Updates added for Rev 26, January 2019 release:
-
Replaced ow2sca8i (Host Software Install Release 9.1) in "Operator Workspace and Computer" with DOC2191008 (9.1 Host Load from Cold Procedure using the repackaged software disk)
-
Renamed Direction 5407677 "MR Magnet Safety Requirements" to "Pre-11.x Service Methods Addendum"
-
Planned Maintenance (Direction 2203808) changed to Rev 9
Updates added for Rev 25, November 30, 2010 release:
- Per iTrak 13196880 and Trackwise 2574982, made the following changes to
the PM Manual - Direction 2203808 to Rev 8:
-
Updated Introduction (Pages
1-4) and included the PM Schedules, thereby obsolescing Directions 2215702
and 2215703
-
Added Section 7-12 to
Chapter 2, PM Procedures
-
Added the Direction number and Rev to all
footers in the A-F PM Schedules
- Per CAPA 1859293, added the new MR Safety Addendum, Direction 5407677, and
updated the following with Warnings/text:
-
ME1REA1.DOC, SRI AND DOCK HARDWARE to Rev 2
-
ME1SCA3.DOC, ENCLOSURE COVERS AND DOCK
ADJUSTMENTS to Rev 4
-
ME2SCA2.DOC, ENCLOSURE COVERS AND DOCK
ADJUSTMENTS FOR OPEN ENCLOSURE to Rev 4
-
MI0REA2.DOC, BLOWER BOX MOTOR REPLACEMENT AND
ADJUSTMENT to Rev 3
-
PT1SCA2.DOC, LIGHTWEIGHT PATIENT TRANSPORT
ADJUSTMENTS to Rev 1
-
PT1SCA4.DOC, DUAL TABLE ALIGNMENT PROCEDURE to
Rev 1
-
RP1REA1.DOC, REAR PEDESTAL HARDWARE REPLACEMENT
FOR NON-CX MAGNET to Rev 3
-
RP1REA1A.DOC, REAR PEDESTAL HARDWARE
REPLACEMENT FOR CX MAGNET to Rev 3
-
RP2REA4.DOC, REAR PEDESTAL REPLACEMENT OPEN
ENCLOSURE to Rev 2
-
PM_SEC1.DOC, Planned Maintenance - Magnet
Room
-
PM_SEC3.DOC, Planned Maintenance - Patient
Handling
-
Direction 2304881, Signa 1.5T TwinSpeed
System Upgrade Installation
-
Direction 2223753, Signa MR/i and CV/i
Upgrades
-
Direction 2248333, Magnet and System Upgrade
to Signa WideOpen (HiSpeed and EchoSpeed)
-
Direction 2223752, Signa MR/i and CV/i
System Installation
-
Direction 2294004, Signa TwinSpeed System
Installation
- Per SPRs MRIhc47186 and MRIhc44718
eliminated the PIN Diode Replacement procedure in:
-
Direction 2135785, Signa Phased Array Shoulder Coil to Rev 2
-
Direction 2131751, Signa 1.5T GP Flex Coil to Rev 3
- Per SPR MRIhc52545, updated Direction 2138247, Signa Illustrated Parts
List
Updates added for Rev 24, November 15, 2009 release:
-
Per SPR SVCge48089: Added a Note to Table 1 and Section 3-2 Configuration for
2292662 Type Medcam Baseboard for 2292662-11.
-
Per PQR 13266277: Added Section 4.1 Using Gradient Insertion Tool 2164744-6 or
Greater.
-
Per PQR 13266277: Added Section 4.1 Using Gradient Insertion Tool
2164744-6 or Greater.
Updates added for Rev 23, June 30, 2009 release:
- Per
PQR 13176796: Created new document named System Errors as a result of FUS
Option.
- Per
PQR 13207656: Updated Direction 2375136-7, 1.5T Mark 9000 Phased Array
Shoulder Coil Service Manual.
- Per
PQR 13212305: Created a new procedure: HP4000 Networked Printer Setup.
- Per
PQR 13220158: Updated Signa Teledyne Surface Coil Test.
- Per
PQR 13198522: Updated Direction 2203808, Section 3: Patient Handling.
Updates added for Rev 22, January 2009 release:
- Per PQR 13112024: Created new document for NEC1990SX and updated Basic or
Generic LCD Monitor Calibration & Gamma Setup.
- Per PQR 13127456: Updated Direction 2205522 to Revision 3 with correct
cardiac phased array coil.
- Per PQR 13171671: Updated Gradient Coil Replacement for BRM, BRM-D &
CRM for use with 4 bolts to Revision 8.
- Per PQR 13165487: Updated MEDCAM AP Board Replacement/Upgrade to Revision
2.
Updates added for Rev 21, June 2008 release:
- Per PQR 13118083: Added Section D-5, Film Artifacts, due to HW issues.
- Per PQR 1106480: Changed Section 1-2 to Safety and moved Warning
statements from former Sections 1-2 and 1-3.
Updates added for Rev 20, March 2008 release:
- Per PSR 13094200 & 13113617: Added EHS Lifting Policy Warning
statements to ACGD SGA Power Supply Replacement.
- Per PSR 13142068: Added Electric Shock Danger statement to Gradient Amp
Power Supply Replacement.
- Per PSR 13118251: Added Geometry Verification Important statement to:
- ACGD Cabinet Troubleshooting Flowchart
- GRAM Troubleshooting Harness
- 8645/GRAM Troubleshooting
- SGD HI Slew Problem Isolation
- SGD Base Cabinet Troubleshooting Flowchart
- Per PSR 13135738: Added Customer Password Management Process.
- Per PQR 13113617: Updated Chiller Vendor Manual to Revision 9.
Updates added since Rev 19 release:
- Oct. 16, 2007: Updated Direction
2223752, Signa MR/i & CV/i:
System Installation Manual to a Revision 14.
- Oct. 17, 2007: Per PSR 13105722, all procedures utilizing the Universal Hoist Kit were
updated with a Warning statement.
- Oct. 17, 2007: Per PSR 13108468, procedures relating to the Cradle Release
were updated.
- Nov. 30, 2007: Revision 4 update for Signa Lx ConnectPro (HIS/RIS) PQR(s):
- 13104792
- 13084868
- 13104789
- 13104791