Recon Performance Verification Procedure
Prerequisites
Overview
This procedure describes how to verify the recon performance speciation which is provided in the Product Datasheet.
1 Acquire Images
Procedure
- Place QA phantom on cradle.
- Set Internal Landmark.
- Select NEW PATIENT.
- Type the following entries in the listed Patient Information
fields:
- Patient ID: Service
- Patient Name: Recon Performance Test
- Select Service from Protocol Selection, Anatomical Selector.
- Select Verify, 42.2 Recon
Performance Test.
The scan protocol is shown as below.
- Click CONFIRM.
- Press MOVE TO SCAN on Scan Control Box when it flashes.
- Press START on Scan Control Box when it flashes.
- By using a stop watch, measure the time between the first image display and the last image display.
2 Calculation of Reconstruction Time
Procedure
- Calculate the reconstruction time (fps) per the following formula:
Reconstruction Time (fps) = (total number of images -1) / (time between the first image display and the last image display)
Specification: ≥50 fps
note: The specification may change by new software release. Confirm with the latest product data sheet.
3 Finalization
No finalization steps.