- Topic ID: id_11038864
- Version: 3.0
- Date: Dec 3, 2019 2:00:11 AM
Parallel Setting of Cradle
Prerequisites
Procedure
- "Remove power from Table by turning off “120VAC”, “Axial Drive” and “HVDC” switches on the service switch panel."
- Remove the Table right and left side covers.
- Move the cradle to the Out Limit position by hand.
- Adjust the position of the bearing to meet the following conditions
(see illustration Figure 1).
-
The Table frame and the cradle are parallel; adjust so that the difference between gap DF (measured on the In side) and gap DR (measured on the Out side) becomes 1.5 mm or less.
-
A gap of 0.1 mm or more (A) exists between the cradle and the bearing (at least one of the right and left bearings) ; however, the total of the gaps on the right and left sides should be 2.5 mm or less.
Figure 1. Gap between Frame and Cradle

-
- Restore the Table to original configuration.
Finalization
No finalization steps.