• Topic ID: id_17423475
  • Version: 3.0
  • Date: Jan 20, 2020 8:32:35 PM

NPSUB Board Assembly Replacement

Prerequisites

Overview

Procedure

    Notice Image
    • notice
    • Understand and follow the Safety Guidelines Manual.
  1. Move the cradle to the OUT mechanical limit position by hand.
  2. Raise the Table to its highest position.
    note:

    If the Table up/down movement is inoperative, use the service power cable to raise the Table (refer to Enforced Table Elevation).

  3. Remove power from Table by turning off “120VAC”, “Axial Drive” and “HVDC” switches on the service switch panel.
  4. Remove the following covers from the Table:

    Figure 1. NPSUB Board Location

  5. Disconnect the three cable connectors (J1 ~ J3) from the NPSUB Board Assy.
  6. Remove the NPSUB Board Assy by unscrewing its four screws.
  7. Install the new NPSUB Board Assy by referring toStep 6 through Step 5.
  8. Restore the Table to original configuration.

Finalization

No finalization steps.