- Topic ID: id_11039096
- Version: 5.0
- Date: Jan 27, 2022 12:51:02 AM
Cradle Replacement
Prerequisites
Procedure
- notice
- Move the cradle OUT to its fully retracted position (farthest from Gantry).
- Raise the Table to its highest position.
- Remove power from Table by turning off “120VAC”, “Axial Drive” and “HVDC” switches on the service switch panel.
- Remove the six Allen screws from the cradle end, and then slide
out the cradle.
Figure 1. Cradle Replacement


caution
- Install the new cradle.
![]() |
Finalization
- After installing the new cradle, perform the following adjustment according to Align, Setup, Cals -> Table.
-
GAP BETWEEN CRADLE AND CRADLE SUPPORT
-