Rotor Diagnostic Test
Overview
Objective: The purpose of this test is to drive the rotation inverter in high speed mode and low speed mode on the tube in order to identify a defective rotation board or defective phase capacitors or a wrong connection among rotation board, HEMIT and tube. Summary: Run diagnostic. The test is executed without high voltage and filament heating, and needs no manual operation.
- Launch the GENERATOR TOOL.
- Launch the Rotor Diagnostic tool.
- Run the test. A rotor sequence will proceed.
- If a problem occurs, check the error code.
1 Finalization
No finalization steps.