- Topic ID: task_gz1_rvx_y3b
- Version: 3.0
- Date: Dec 29, 2020 1:00:01 AM
Configure HIPAA procedure
Prerequisites
Overview
This document describes how to enable Data Privacy (User Authorization) feature on CT system.
This procedure applies to the following system:
- Optima CT620 (19BWxx.xx)
Procedure
- Click the Service Icon on the Display Monitor, and the Common Service Desktop (CSD) will appear.
- Select the Configuration Tab on the CSD and find the Configure HIPAA selection in the menu.
Figure 1. Common Service Desktop

- Login as EA3 admin user. Ask Customer’s system administrator to login.
- In the next screen, select HIPAA ON and click SUBMIT.
Figure 2. HIPAA Selection

- Reboot the system to for HIPAA configuration change to take effect.
Finalization
No finalization steps.